• Germany
  • Austria
  • Italy
  • Russia
  • Czech Republic
  • Slovenia
  • Poland
  • Turkey

MGA5+

PORTABLE, NDIR EMISSION MONITORING SYSTEM. High accuracy emission analysis.

High-quality, portable INFRARED MULTIGAS ANALYZER
for emission monitoring and combustion analysis

  • Double stage gas cooler with automatic condensate draining pump
  • Heated gas sampling line with lengths of 9.84′ / 16.4′ / 32.81′ or 65.62′
  • Automatic internal test and control of soft and hardware functions
  • Large, high contrast and backlit graphical display with zoom function
  • RS 232 interface and internal data memory for approx. 8,500 measurements
  • RS 485 interface for external MRU smart sensor (transmitter) connection
  • Automatic interval measurement
  • Data visualization and evaluation software for PC (32bit Data Logger)
  • Solenoid valve for automatic zeroing plus calibration
  • Universal analog input (4 … 20 mA or 0 … 10 V) or additional NiCrNi thermocouple input
  • 8 channel analog output 4 … 20 mA with user configurable output
  • Automatic calibration using integrated calibration gas cells, without use the of external calibration gas cylinders
Measuring ranges
O2 – Oxygen 0 … 25.0 % (Electro-chemical or Paramagnetic)
CO2 – Carbon dioxide 0 … 4% / 20% (min / max range) (NDIR)
CO – Carbon monoxide 0 … 200ppm / 1,000 ppm (min / max range) (NDIR)
NO – Nitric oxide 0 … 200ppm / 1,000 ppm (min / max range) (NDIR)
NO2 – Nitrogen dioxide Catalytic converter NO2 to NO min. 90 % conversion efficiency
SO2 – Sulfur dioxide 0 … 200ppm / 1,000 ppm (min / max range) (NDIR)
CH4 – Methane 0 … 200ppm / 1,000 ppm (min / max range) (NDIR)